C O M M A N D R E F E R E N C E
Command Reference
04_09
LDC-3700C Series
117
C H A P T E R
4
The TEC:ENABle:COND? query returns the value of the status condition enable register
of the TEC operations.
Notes
The enabled TEC conditions can be set by using the TEC:ENABle:COND command.
The enabled TEC condition status can be monitored by the TEC:COND? query.
Examples
"TEC:ENAB:COND?" -response: 129, means that the TE Module Open and TE Current
Limit conditions may be reported in the status byte register.
"Radix Hex; TEC:Enable:Cond?" -response: #H0EFB, means that any and all of the
above conditions will be reported in the status byte register.
TEC:ENABle:COND?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
1 - TE Current Limit
256 - N/A
2 - Voltage Limit Error
512 - Output Out of Tolerance
4 - N/A
1024 - Output On
8 - High Temperature Limit
2048 - Ready for Calibration Data
16 - TEC Interlock Enable
4096 - N/A
32 - Booster Enable
8192 - N/A
64 - Sensor Open
16384 - N/A
128 - TE Module Open
32768 - N/A