C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Calibration Overview
04_09
LDC-3700C Series
145
C H A P T E R
5
6
Release the (
PARAMETER
)
SET
switch to store the first calibration value into non-volatile
memory. The controller will then set the second calibration current, approximately one-
fourth (1/4) of the original current. (For example, if the first calibration set point was 4000
µA, the second set point will be about 1000 µA.)
7
Press and hold in the (
PARAMETER
)
SET
switch and turn the
ADJUST
knob until the
LASER display indicates the actual current as measured by the ammeter or as calculated.
8
Release the (
PARAMETER
)
SET
switch to accept the second calibration point. After the
(
PARAMETER
)
SET
switch is released, the LDC-3700C Series Laser Diode Controller
will calculate the calibration constants and store them to nonvolatile memory.