C O M M A N D R E F E R E N C E
Command Reference
04_09
LDC-3700C Series
127
C H A P T E R
4
The TEC:OUTput? query returns the status of the TEC output switch.
Notes
Although the status of the switch is on, the output may not have reached the set point
value.
Examples
"Tec:Out?" -response: 0, means the TEC output switch is off, TEC output is off.
"TEC:OUT?" -response: 1, means the TEC output switch is on, TEC output is on.
The TEC:R command sets the TEC's constant thermistor resistance or linear sensor
reference set point.
Parameters
An <nrf value> which represents the thermistor resistance set point value, in k
; the
AD590 current set point, in
A; or the LM335 voltage set point, in mV, depending on the
selected sensor type.
Note
The R set point is used to control the TEC output in R mode only. Using the R mode, the
user may also monitor the temperature of the TEC load via a remote algorithm of his/her
own design.
Examples
"TEC:R 20.5" -action: sets the set point thermistor resistance to 20.5 k
, or sets the
AD590 current to 20.5
A, or sets the LM335 voltage to 20.5 mV.
"TEC:Mode:R; TEC:R 10" -action: sets the set point thermistor resistance to 10.0 k
, or
the AD590 current to 10.0
A, or the LM335 voltage to 10.0 mV. The TEC output sensor
is controlled to that value, if the output is on.
The TEC:R? query returns the value of the TEC thermistor resistance, or AD590 current,
or LM335 voltage.
Notes
TEC load temperature is derived from the thermistor resistance or linear sensor
reference measurement.
This measurement is updated approximately once every 600 msec.
Examples
"TEC:R?" -response: 10.543, means the measured TEC thermistor resistance is 10.543
k
, or the measured AD590 current is 10.543
A, or the measured LM335 voltage is
10.543 mV, depending on the SENSOR SELECT switch position.
"Tec:R?" -response: 0.728, means the measured TEC thermistor resistance is 728
ohms, or the measured AD590 current is 0.728
A, or the measured LM335 voltage is
0.728 mV, depending on the SENSOR SELECT switch position.
TEC:OUTput?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
TEC:R
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
TEC:R?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Summary of Contents for LDC-3700C Series
Page 2: ......
Page 8: ...L I S T O F FIG U R E S vi LDC 3700C Series...
Page 10: ...L I S T O F TA B L E S viii LDC 3700C Series...
Page 78: ...RE M O T E O PE R A T I ON Command Syntax 62 LDC 3700C Series C H A P T E R 3...
Page 150: ...C O M M A N D RE F E R E N C E Command Reference 134 LDC 3700C Series C H A P T E R 4...