O P E R A T I O N S
Operating the TE Controller from the Front Panel
34
LDC-3700C Series
C H A P T E R
2
switch is released, the new value is stored in non-volatile memory and becomes
the maximum set point value.
Adjusting the Temperature Set Point
Now, you are ready to adjust the set point operating temperature of the controller
to 35°C. The LDC-3700C Series Temperature Controller will monitor actual
temperature and adjust TE current to maintain the thermal load at the set point
temperature.
To adjust the Temperature Set Point to 35°C,
set the instrument to
TEC
mode by
pressing the
TEC
push button in the
ADJUST
section. This instrument mode will
be indicated by the
TEC
LED. Turn the adjust knob clockwise to increase the set
point (as indicated on the
TEC
display), or counterclockwise to decrease the set
point. When the adjust knob is turned, the controller automatically changes to set
point mode indicated by the display enunciator flashing. Before adjusting the set
point check the
TEC
display for proper display mode. The display mode
determines which set point is being adjusted. Adjust the set point temperature to
35°C. Three seconds after the set point is adjusted to the new value, the
instrument will return to measurement mode. Pressing the
SET
switch in the
TEC
display switch section will also change the controller to set point mode and display
the set point corresponding to the display mode.
Once the sensors, temperature control loop parameters and set points are
established, the output of the temperature controller can be enabled. With the
output enabled, as indicated by the
ON
LED, the temperature controller will drive
current to the TE modules to cool or heat the load depending on the difference
between the set point and the actual temperature. Press the
OUTPUT
switch
once to enable the
TEC
output, and again to disable the output. The default
condition for the
TEC
output is off when the instrument is first powered up.
TEC
LAS
ADJUST
TEC MODE
ON
R
I
TE
T
OUTPUT
SELECT
Summary of Contents for LDC-3700C Series
Page 2: ......
Page 8: ...L I S T O F FIG U R E S vi LDC 3700C Series...
Page 10: ...L I S T O F TA B L E S viii LDC 3700C Series...
Page 78: ...RE M O T E O PE R A T I ON Command Syntax 62 LDC 3700C Series C H A P T E R 3...
Page 150: ...C O M M A N D RE F E R E N C E Command Reference 134 LDC 3700C Series C H A P T E R 4...