In order to be able to get the EXTRACTION option, the freeze time should have been at
least 2 minutes.
If you did not previously freeze for 2 minutes, the system may update you to wait for a
passive thawing process (fig. 85) before you can extract the cryoprobe.
•
To activate the EXTRACTION, press on the optional RED button on the straight
cryohandle or the EXTRACTION icon on screen as previously shown.
•
The EXTRACTION screen will be displayed and a time count will appear.
•
At the end of the EXTRACTION step, a message will be displayed on screen
informing the user that the EXTRACTION cycle is done (for example 40-50 seconds
for 2 cm ice ball).
•
Wait for the message, then gently remove the cryoprobe from the target tissue and
then press FINISH. Do not force removal of the cryoprobe from the tissue as it
might increase the risk of hematoma.
•
If the cryoprobe cannot easily be extracted from the tissue, press the RED HANDLE
button on the straight cryohandle or EXTRACTION icon on screen to initiate another
EXTRACTION cycle.
•
After the cryoprobe has been removed from the tissue, press the Action button on
the straight
cryohandle or the “Finish” icon on screen to complete the procedure.
•
After the cryoprobe has been removed, apply momentary pressure to the insertion
site. You may apply adhesive skin closure to the incision and cover it with a 4x4
gauze dressing.