Figure 68: Cryoprobe expired
Warning
Never reuse a single-use cryoprobe, temperature sensor or a single-use
sterile sleeve.
Warning
Never “unscrew” the cryoprobe if you are not clearly
instructed to unscrew
or disengage it.
Warning
After completion of treatment of a given patient, make sure that the single-
use cryoprobe and single-use sterile sleeve are removed and safely discarded.
Warning
Before beginning treatment of a patient, you MUST ensure that the previous
single-use cryoprobe and sterile sleeve have been removed.
Warning
Cryoprobes are fragile and can be damaged if mishandled. Do not use a
cryoprobe that has been bent, dropped, hit against a hard surface or
compromised in any manner, as damage to the cryoprobe may have
occurred.
Warning
1. While cryoprobe is in use, surgical instruments should be handled with
care to avoid damage to the cryoprobe.
2. Damaged cryoprobe can become inoperable. Thus, avoid pinching, cutting
or pulling it excessively.