Follow system instructions until you are required to safely remove the cryoprobe
from the cryohandle.
•
You will then be returned to the Main Menu screen
.
Check that the Dewar is full
and replace the cryoprobe and start again. If the problem recurs do not proceed
with treatment. Turn OFF the system and contact IceCure Medical.
•
Choosing SUCCESS will load the treatment selection screen.
•
If for any other reason you decide to stop the procedure, press CANCEL. Follow
system instructions until you are required to safely remove the cryoprobe from
the cryohandle. You will then be returned to the Main Menu screen.
•
Do not try to disengage a cryoprobe before instructed since the system is under
pressure and you can be hurt by the sharp cryoprobe being ejected under
pressure.
•
In addition, the hose and the cryoprobe still contain liquid nitrogen that can be
spilled out if you disengage the cryoprobe before you get proper system
instructions to disengage the cryoprobe.
Warning
In case of frost on shaft, start EXTRACTION if possible. If not, wait for passive
THAW. In both cases use skin protection techniques.
Warning
During pre-test, portions of the cryoprobe other than the freeze zone, including
the plastic cover that is located near the cryoprobe handle, may become cold
due to malfunction. If unwanted freezing occurs, stop the pre -test process by
pressing CANCEL.
Warning
Pay attention to the cryoprobe during pre-test: bubbles, leak or frost on shaft
may signal internal
problems that could be hazardous.
5.2.7
Treatment Selection
The system allows you to choose between manual mode and automatic mode using
preset protocols.
To choose automatic mode, select one of the preset protocols from the Treatment
Selection screen as shown below.