DSR3200000 rev. G
ProSense
™
Confidential
User Manual
75
Return to first page U.S User Manual
5.2.6.6
Cryoprobe connection
Connect the cryoprobe to the cryohandle as follows, while maintaining sterility of the
cryoprobe:
•
Open the cryoprobe pouch in accordance to the “PEEL
HERE” label
•
Remove the plug that covers the cryoprobe connection
point.
•
Insert the cryoprobe into the insertion point in the handle
as shown on screen and screw it until a NEXT button
appears on screen then confirm screwing by an additional
slight rotation to confirm that cryoprobe connection is
secured.
•
Remove the cryoprobe cover that protects the tip.
Figure 64
: “PEEL HERE” label
•
When done, press the Action button on the handle or NEXT on the touch screen.
•
After pressing NEXT
, a “Do not disconnect” message will appear
Figure 65: Cryoprobe connection screen
•
Cryoprobe reading failed
message appears if the cryoprobe reading failed. Make
sure the cryoprobe is secured and press OK to cryoprobe rereading.