
IceCure
™
Medical Ltd.
DSR3200000 rev. G
ProSense
™
Confidential
User Manual
137
Return to first page U.S User Manual
Recommended separation distances between portable and mobile RF communications
equipment and the EQUIPMENT or SYSTEM- for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and
the model
ProSense™
Cryotherapy product
The model
ProSense™
Cryotherapy product is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the model
ProSense™
Cryotherapy product can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the model
ProSense™
Cryotherapy product as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.