3.3.1
Main chassis
The
ProSense™
cryoablation system is housed within a chassis mounted ON four rollers
for ease of movement. Each roller is equipped with directional and rotational brakes for
system immobilization. Located ON the top of the chassis are a touch screen control
panel and a cryohandle cradle. On the right upper part of the chassis is the Emergency
Stop button - a round red button that shuts down the system immediately in an
emergency situation. On the back of the chassis are two hooks used for hanging the
electric cable, and a grip handle for ease of system transportation.
Figure 23: The
ProSense™
transportation rollers & brakes
3.3.2
Emergency Stop button
The Emergency Stop button is a red knob located ON the right upper side of the chassis.
It is designed for emergency shutdown of the unit. Pressing this button immediately
turns OFF the electrical power supply to the system. To release the Emergency Stop
button, turn it in the direction of the arrows.
Figure 24: The Emergency Stop button
Caution
Only push the Emergency Stop button when there is NO other choice.
Whenever possible, use EXTRACTION to release the cryoprobe from the
tissue and use the standard shutdown procedure. Excessive use of the
Emergency Stop button may damage the system
.
Warning