INSTALLATION AND SETUP
4.1
Space and positioning requirements
The work area for the unit should be prepared as per the dimensions described in the
system specifications (chapter 11). In order to guarantee sufficient ventilation, always
maintain a clearance distance of at least 0.5 meters (20 inches) between the unit and
walls or other objects that may obstruct air flow.
Adequate ventilation and air circulation are major considerations when working with
liquid nitrogen.
Warning
Liquid nitrogen may cause
, if there is not sufficient ventilation in the room,
serious injury or burn if handled improperly. Local laws and safety rules
regarding the maintenance and handling of liquid nitrogen Dewars should
always be observed.
4.2
Setup warnings and cautions
Warning
If there is not sufficient ventilation in the room, the
ProSense™ cryoablation
system cannot be used due to risk of suffocation due to increased levels of
nitrogen in the room.
Caution
The ProSense
™ cryoablation system must be unpacked, installed, and tested
by an IceCure Medical authorized technician only.
Caution
After positioning the main chassis, lock the front roller brakes. Failure to do
so may result in damage to the system or to other equipment in the clinic
room.
Caution
Do not move the system when the Dewar contains liquid nitrogen.