TROUBLESHOOTING
9.1
General
Warning
Do not modify this equipment without authorization of the manufacturer.
Warning
Never open the console. Only an IceCure Medical technician or authorized
representative is allowed to open the console for maintenance or to
repair the system.
Warning
Do not attempt to perform any troubleshooting or corrective action
beyond those specified in the following guide. Any malfunction not listed
in the guide, or one that persists after the recommended action has been
taken, must be referred to IceCure Medical
.
9.2
Troubleshooting guide
Problem
Probable Cause
Action
Main chassis does not
move
Wheels are locked
Unlock the wheels, transport the
system and lock them again.
If the problem persists, contact
IceCure Medical service.
No system power
AC power is not connected to the
system; power supply malfunction
Check that the power cable is
connected to the inlet and the wall
outlet.
Check that Emergency button is
released.
Check that main switch is ON.
If still there is no power, contact
IceCure Medical service.
AC power is ON but the
screen does not turn ON
Computer power supply
malfunction or computer error
Check if the computer power cable
is connected to the screen and
turn the screen ON.
If the screen still does not turn ON
contact IceCure Medical service.
The system and screen
turn ON
but doesn’t
enter to the Main screen
AC power supply malfunction
Contact IceCure Medical service