User Manual
105
Return to first page U.S User Manual
Figure 104: Replacement was completed after the THAW ended
5.3.8.1
Replace the Dewar during Freeze
In case of low Nitrogen,
the system can’t continue to Freeze.
It will ask to refill the
Dewar in Freeze cycle (see fig. below)
Figure 105: Replacement during Freeze, in case of empty Dewar
.
•
If you press OK, you can replace a Dewar and continue the procedure. The system will
move to Manual mode and wait in a THAW cycle for instruction.
Pay attention that in Manual mode, you need to press the icon on the screen for each
freeze or THAW cycle.
•
Replace Dewar as in paragraph 5.3.8.