Return to first page U.S User Manual
Problem
Probable Cause
Action
working/ Controller is
stuck
the back of the system or
emergency stop.
Then turn ON the system (after
releasing the emergency stop
button, if pressed) and wait for
computer re-boot.
Interference in adjacent
devices such as image
artifacts on screen
Unknown
First make sure the problem is
related to the system by examine
the interference after turning the
system off.
If the interference is still observed:
▪
Re-orient or relocate the
other device(s).
▪
Increase the separation
between the equipment.
▪
Connect the system into an
outlet on a circuit different
from that to which the other
device(s) are connected.
If the problem persists, contact
IceCure Medical service.
“Long wash” popup
message
System pipes blockage
The user will have to wait until the
wash ends.
In order to have better indication
in regards to the end of the long
wash procedure, it’s better not to
click “O.K” when the message
pops up, but to wait until the
message disappear by itself. In
case you clicked “O.K” the screen
freeze till the end of the “long
wash” procedure.
Table 2: Troubleshooting guide