Return to first page U.S User Manual
Warning
After extraction step, before removing the cryoprobe from the tissue, make
sure that the freeze effect has been deactivated so that the cryoprobe can
easily be removed from the tissue.
Do not force removal of the cryoprobe from the tissue as it might increase
the risk of tissue damage. Continue the Extraction step or wait for passive
thaw until the cryoprobe can be withdrawn easily.
Warning
If one or more of the wheels is damaged, do not use the system.
Warning
Cryoprobes are fragile and can be damaged if mishandled. Do not use a
cryoprobe that has been bent, dropped, hit against a hard surface or
compromised in any manner, as damage to the cryoprobe may have
occurred.
Warning
1. while cryoprobe is in use, surgical instruments should be handled with
care to avoid damage to the cryoprobe.
2.Damaged cryoprobe can become inoperable. Thus, avoid pinching, cutting
or pulling it excessively.
3.Do not grasp cryoprobes with auxiliary instruments as this may cause
damage to the cryoprobe shaft.
Caution
Do not move the system when the Dewar contains liquid nitrogen.
Warning
Removing the Dewar, or placing it back within the system after refilling it
must ONLY be done according to system instructions and with the carriage in
the bottom position. If the carriage is not in the bottom position, liquid
nitrogen may spill out.
Caution