Table 8: Channel numbers are dedicated as follows:
Channel number
Parameter
Channel 1
Flow (outlet gas flow rate)
Channel 2
CO
2
Channel 3
Humidity
Channel 4
T-C (Humidifying Column Temperature)
Channel 5
O
2
*
*CO
2
/O
2
version only
Alarms
In this dialog it is possible to set the alarm limits for divergent control parameters. Move
from the mode menu to the alarm menu by pressing the ”right” button.
The maximum and minimum limits of the alarm can be set for each individual channel. If the alarm is
activated (which happens when the current value goes under/over the low/high limits), the current
value (I) blinks, showing alternately the value and ”low” or ”high”, respectively. To stop the display
blinking, navigate to the blinking channel and press the round button once.
<Alarms>
Setup
Channel = 4
High = 42°C
Low = 36°C
Figure 22: Alarm settings dialog.
Preferences
Set the brightness and contrast of the display in the preferences menu.
<Preferences>
Setup
Backlight = 65
Contrast = 15
Return
Figure 23: Preferences dialog.
Info
Information about serial number and firmware version is shown in the info dialog.
Figure 24: Info display.
Version 1.0 (June 1, 2022)