ibidi Gas Incubation System – Silver Line
5.4
Setting Parameters with the Front Display
All control parameters can be manually set on the controller using the buttons and the display on the
front panel.
ibidi Gas Mixer CO
2
/ O
2
Flow
I 9.9
S[ 10.0]
Run
CO2% Hum%
5.1
5.0
75.2
80.0
Setup
Up/right
Down/left
Round button
Figure 18: Gas Mixer front display and set buttons.
The display shows the measured (’I’ = instantaneous) and set (’S’ = set) temperatures of all the chan-
nels.
The cursor position is indicated with square brackets (“
[
]
”). You can move the cursor using the
“left” and “right” buttons. If you want to select a parameter or a function, press the round button
and the square brackets will change to angle brackets (“
<
>
”). Now you are able to change the
value with the “up” and “down” buttons. To confirm the changed value, you must press the round
button once more.
5.4.1
Run/Stop Mode
Set the whole system to run or stop mode by manipulating the setting in the display’s bottom line to
“Run” (= system is running) or “Stop” (= system is not running).
Figure 19: Main display indicating “Run”.
Instruction Manual
31
Version 1.0 (June 1, 2022)