ibidi Gas Incubation System – Silver Line
CAUTION
• Ensure that the external power supply is easily accessible. The ibidi Gas Incubation System –
Silver Line must be installed in a manner such that none of its components hinders access to
the external power supply.
• Immediately replace damaged cords, plugs, or cables to avoid risk of personal injury or damage
to the instrument.
• Only ibidi technical staff and technical staff instructed by ibidi are permitted to open and ser-
vice the ibidi Gas Incubation System – Silver Line.
• The external power supply should not be brought into contact with moisture. If the housing is
damaged, the external power supply should not be used.
• Avoid strong magnetic fields and sources of high frequency. The ibidi Gas Incubation System
– Silver Line might not function properly when located near a strong magnetic field or high
frequency source.
• Avoid vibrations from vacuum pumps, centrifuges, electric motors, processing equipment, and
machine tools.
• Avoid dust and corrosive gas. Do not install the ibidi Gas Incubation System – Silver Line
where it could be exposed to high levels of dust or to outside air or ventilation outlets.
• Install the ibidi Gas Incubation System – Silver Line in a location that enables easy access for
maintenance.
• Do not place heavy objects on the instrument.
• Install the ibidi Gas Incubation System – Silver Line in a horizontal and stable position, such as
a table, bench, or desk upon which the instrument is installed.
• The Gas Mixer needs a gas input of 1 bar/14.5 psi. Only operate the Gas Mixer with the pressure
indicated in the specifications on page
Instruction Manual
11
Version 1.0 (June 1, 2022)