Step 5: Adjust the Image Quality
Related Setting Items:
Note:
For the details on the short ASCII commands required to configure the image quality
settings, see "
".
To maximize the performance of the camera, configure its basic function in the following order.
1. Configure the Line Rate.
For more information, see
.
2. Configure the Exposure Time.
For more information, see
.
Note:
The line rates and exposure times that can be set will vary depending on the
CameraLinkClockFrequency and ClConfiguration settings for Camera Link.
3. Perform DSNU and PRNU corrections.
For more information, see
.
4. Adjust the Black Level.
For more information, see
.
5. Adjust the White Balance using the automatic adjustment function.
a. Place a white sheet of paper or similar object under the same lighting conditions as
the intended subject, and zoom in to capture the white.
White objects near the subject, such as a white cloth or wall, can also be used. Be
sure to prevent the high-intensity spot lights from entering the screen. The white
balance is automatically adjusted.
b. Select the
BalanceWhiteAuto
tab, and select
Once
.
The white balance is automatically adjusted.
Step 6: Configure Various Other Settings
See “
to configure settings as necessary.
Note:
We recommend performing DSNU and PRNU calibration again whenever the line rate