B-28
i-CAT FLX Technical Guide
032-0328-EN Rev K
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The
i-CAT FLX
is intended for use in the electromagnetic environment. The customer or the user of the
i-CAT FLX
should ensure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment – Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/-6 kV contact
+/-8 kV air
+/-2, 4 & 6 kV
contact
+/-2, 4 & 8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/-2 kV for power
supply lines
+/-1 kV for input/
output lines
+/-2 kV for power
supply lines
+/-1 kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+/-1 kV line(s) to
line(s)
+/-2 kV line(s) to
earth
+/-0.5 & 1 kV
differential mode
+/-0.5, 1 & 2 kV
common mode
Mains power quality should be
that of a typical commercial or
hospital environment.