3-14
i-CAT FLX Technical Guide
032-0328-EN Rev K
QA Air Water Test
Set Up QA Air Water Phantom
NOTE:
It is important to use the water phantom provided with the scanner. Use distilled
water in the phantom. Using tap water may negatively affect the test results.
1.
Remove chin cup and insert phantom platform at lowest position.
2.
Fill phantom half full with distilled water and carefully
place on platform.
3.
Using the Alignment Lasers, center the water bath
with the horizontal laser across the center of the
water depth.
Run QA Air Water Test
1.
From Utilities menu, select
QA Air Water
.
2.
Ensure phantom is set up properly, then press
.
3.
Select
, then press
. The scanner initializes.
4.
When prompted, press the
Scan
button on the operator control box. An audible alarm is
sounded and the X-ray ON light is illuminated during radiation exposure.
5.
Review scout image. The phantom must be centered. Adjust the phantom platform as
needed to achieve the proper height.
6.
To move the phantom to the right or left, use the
Front/Back
slider control.
If required, make adjustments, then press
to run
again. Repeat as
required until phantom is properly aligned.
7.
When phantom is aligned, select
and press
.
8.
When prompted, press the
Scan
button on the operator control box. An audible alarm is
sounded and the X-ray ON light is illuminated during radiation exposure.
Summary of Contents for FLX V Series
Page 36: ...3 24 i CAT FLX Technical Guide 032 0328 EN Rev K ...
Page 60: ...A 8 i CAT FLX Technical Guide 032 0328 EN Rev K ...
Page 78: ...B 18 i CAT FLX Technical Guide 032 0328 EN Rev K Scanner Dimensions TOP VIEW FRONT VIEW ...
Page 120: ...D 8 i CAT FLX Technical Guide 032 0328 EN Rev K ...
Page 121: ......