3-6
i-CAT FLX Technical Guide
032-0328-EN Rev K
Make sure the phantom is centered left to right and front to back. Use the lasers to
confirm.
Run QA Line Pair Test
1.
From Utilities menu, select
QA Line Pair
.
2.
Ensure phantom is set up properly, then press
.
3.
Select
, then press
. The scanner initializes.
4.
When prompted, press the
Scan
button on the operator control box. An audible alarm is
sounded and the X-ray ON light is illuminated during radiation exposure.
5.
Review the scout image. The phantom must be centered and level. Adjust the phantom
platform as needed to achieve the proper height.
6.
To move the phantom to the right or left, use the
Front/Back
slider control.
If required, make adjustments, then press
to run
again. Repeat as
required until phantom is properly aligned.
7.
When phantom is aligned, select
and press
.
Horizontal Laser Line
through Center of Phantom