3-9
Calibrations and Quality Assurance
032-0328-EN Rev K
9.
Verify that definition is present within line pairs 10, 11, and 12. Compare image quality to
the image shown.
Distance Measurement Test
To ensure measurement accuracy, this procedure checks Distance measurements.
1.
Select
to activate Distance tool.
NOTE:
To better view image, select
to turn off cursor lines.
2.
Click on the outside line of set 16, and then click on the outside line of set 10 to draw a
line, as shown.
3.
The measurement should be between 41 and 42 mm.
4.
When finished, save the file and close TxSTUDIO.
11
10
12
13
14
15
16
Distance Line
Summary of Contents for FLX V Series