C-15
Radiation Information
032-0328-EN Rev K
X-ray Tube Data
X-ray Tube Type: SXR 130-15-0.5
Available focal spot size:
0.5
Available target angle: 15°
Anode construction: Vacuum cast copper
with tungsten target
Cathode construction: Vacuum tube nickel
with tungsten filament
Max. Tube operating voltage: 130 kVp Full
wave rectified
Inherent Filtration: 1.1 mm Al equivalent /
Glass
Cooling Method: High dielectric transformer
oil
Conditioning of Tube Head:
Must be at or above 22
o
C before first scan.
Allow scanner to acclimatize for two hours if
operating environment was out of range or
scanner was stored out of range.
X-Ray Tube Voltage
X-Ray Tube Current
Electrical Power
90 kV
3mA
270W
84 kV
5mA
420W
94 kV
5mA
470W
120 kV
5mA
600W
SXR-130-15-0.5