3-7
Calibrations and Quality Assurance
032-0328-EN Rev K
8.
When prompted, press the
Scan
button on the operator control box. An audible alarm is
sounded and the X-ray ON light is illuminated during radiation exposure.
The scanner acquires data and a status indicator shows acquisition and image creation
progress. When image processing is complete, image is displayed.
9.
Review image to ensure adequate quality. Press
to display Complete screen and
select option to go Back to Utility.
QA Line Pair Evaluation
1. At a clinical workstation, start SmartScan STUDIO Manager and select
Exam List
.
2.
On Exam List, locate QA 3D_1 Line Pair exam with most current date. Double-click CT
entry to load study in TxSTUDIO.
NOTE:
If a pop-up message displays stating “Tru-Pan failed to process”, click
OK
and
continue.
3.
When exam is loaded, select
Section
tab at top of display.
4.
In the upper right corner view, click where the vertical and horizontal cursor lines cross
and drag to the center of the line pairs, as shown.