8-5
Calibration and Maintenance
032-0329-EN Rev K
Routine Cleaning
Routinely clean and disinfect all items which come in contact with the patient, using one of the
products listed, or equivalent cleaner and disinfectant:
Opti-Cide
3®
Solution and/or Wipes from Biotrol International
CaviCide
®
from Metrex (DIN:02161656)
Saturate with disinfectant and allow surface to remain wet for three minutes at room temperature
(69
o
F / 20
o
C), and then wipe dry using a clean paper or cloth towel. See the label of the
disinfectant for full instructions.
The following items may come in physical contact with the patient during a scan:
•
Patient Emergency Stop
•
Head Support
•
Chin Cup/Chin Rest
•
PAN Head Holder
•
Booster Seat/Foot Stool
•
Bite Tips
Clean the equipment surfaces frequently, especially if corroding chemicals are present. Unless
otherwise instructed, use a cloth moistened with warm water and mild soap and wipe all surfaces
to remove surface dirt and marks. Do not use strong cleaners and solvents as these may damage
the finish. It is recommended to power off equipment before cleaning. Be careful to avoid liquid
leaking inside the gantry.
Summary of Contents for FLX V Series
Page 6: ...032 0329 EN Rev K vi i CAT FLX User Manual...
Page 12: ...1 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 20: ...3 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 28: ...4 8 i CAT FLX User Manual 032 0329 EN Rev K...
Page 47: ...6 9 Taking a Scan 032 0329 EN Rev K Choose a Protocol V17 Option V10 Option V8 Option...
Page 63: ...6 25 Taking a Scan 032 0329 EN Rev K Choose Anatomy V17 Option V10 Option V8 Option...
Page 76: ...7 8 i CAT FLX User Manual 032 0329 EN Rev K...
Page 82: ...8 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 84: ...9 2 i CAT FLX User Manual 032 0329 EN Rev K...
Page 94: ...B 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 95: ......