Positioning the Patient
032-0329-EN Rev K
Scouts and Dry Runs
NOTE:
Scout scans and Dry Runs are described in detail in the
Taking a Scan
chapter.
It is recommended that a Scout scan be performed to verify patient positioning. A Scout scan
exposes the patient to a fraction of a second of radiation, but provides an image that can be used
to verify that the patient is positioned to capture desired field of view (FOV).
A Dry Run is recommended for patients who want a demonstration of the machine motion, and
to check shoulder clearance. A Dry Run simply rotates the gantry in the exact manner that is
required for the selected protocol, but does not expose the patient to any radiation.
Occlusal Plane Flat,
Frankfurt Plane tilted
up slightly.
Position of Alignment Lights:
Horizontal:
Occlusal plane between the lips.
Vertical:
1.5 inches in front of condyle (chin support adjusted to acquisition position).
Summary of Contents for FLX V Series
Page 6: ...032 0329 EN Rev K vi i CAT FLX User Manual...
Page 12: ...1 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 20: ...3 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 28: ...4 8 i CAT FLX User Manual 032 0329 EN Rev K...
Page 47: ...6 9 Taking a Scan 032 0329 EN Rev K Choose a Protocol V17 Option V10 Option V8 Option...
Page 63: ...6 25 Taking a Scan 032 0329 EN Rev K Choose Anatomy V17 Option V10 Option V8 Option...
Page 76: ...7 8 i CAT FLX User Manual 032 0329 EN Rev K...
Page 82: ...8 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 84: ...9 2 i CAT FLX User Manual 032 0329 EN Rev K...
Page 94: ...B 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 95: ......