4-5
Selecting a Protocol
032-0329-EN Rev K
NOTE:
•
8 x 5 and 8 x 8 scans capture a smaller field of view. Use a larger volume size when the
anatomical area being scanned is greater than 8 x 5 or 8 x 8.
Panoramic
Standard
94 / 5
20.0
146.4
Panoramic - small
Standard
84 / 5
18.3
91
i-CAT FLX V8 Favorites
CT Favorite Name
Volume Size
(Diameter x Height)
Voxel
(mm)
Scan Time
(Seconds)
DAP
(mGy
cm
2
)
Arches - small
8 x 8
0.3
8.9
239
Arches - small Quick Scan
(Low Dose)
8 x 8
0.3
4.8
134.8
Arches - small Quick Scan+
(Ultra Low Dose)
8 x 8
0.6
4.8, reduced mA
40
Arches - small HD
8 x 8
0.2
26.9
500.8
Arches - small UltraHD
8 x 8
0.125
26.9
500.8
Mandible - small
8 x 5
biased for mandible
0.3
8.9
168
Maxilla - small
8 x 5
biased for maxilla
0.3
8.9
168
Mandible - small UltraHD
8 x 5
biased for mandible
0.125
26.9
341
Maxilla - small UltraHD
8 x 5
biased for maxilla
0.125
26.9
341
PAN Favorite Name
Projection
kVp /
mA
Scan Time
(Seconds)
DAP
(mGy
cm
2
)
Panoramic - enhanced
Standard
94 / 5
20.0
146.4
Panoramic - small enhanced
Standard
84 / 5
18.3
91
Panoramic
Standard
94 / 5
20.0
146.4
Panoramic - small
Standard
84 / 5
18.3
91
i-CAT FLX V10 Favorites
CT Favorite Name
Volume Size
(Diameter x Height)
Voxel
(mm)
Scan Time
(Seconds)
DAP
(mGy
cm
2
)
Summary of Contents for FLX V Series
Page 6: ...032 0329 EN Rev K vi i CAT FLX User Manual...
Page 12: ...1 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 20: ...3 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 28: ...4 8 i CAT FLX User Manual 032 0329 EN Rev K...
Page 47: ...6 9 Taking a Scan 032 0329 EN Rev K Choose a Protocol V17 Option V10 Option V8 Option...
Page 63: ...6 25 Taking a Scan 032 0329 EN Rev K Choose Anatomy V17 Option V10 Option V8 Option...
Page 76: ...7 8 i CAT FLX User Manual 032 0329 EN Rev K...
Page 82: ...8 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 84: ...9 2 i CAT FLX User Manual 032 0329 EN Rev K...
Page 94: ...B 6 i CAT FLX User Manual 032 0329 EN Rev K...
Page 95: ......