Status messages – CS1000 (optionally with display)
OLFCM 15/30/45/60
en(us)
65 / 84
BeWa OLFCM 15-60 3595923a en-us 2018-06-28 .docx
2018-06-28
Error
LED
Flash code /
display
Status
To do
Error
no.
Analogue output
Switching output
Flow status on the
digital output
Red
It is not possible to
determine the flow
rate.
The sensor status
is undefined.
Check that the flow
is between 30 …
500 ml/min.
If the fluid
cleanliness is below
the measurement
limit (ISO 9/8/7), it
may take several
measurement
cycles until
measured values
are displayed.
3
- mA / - V
open
-1
Red
The sensor is
above its
measurement
range
ISO 25/24/23.
It is not possible to
determine the flow
rate.
Filter the fluid.
3
19.9 mA / 9.95 V*
open
-1
CHECK
2%2$2§
2DIRTY