Status messages – SMU1200 (optional)
OLFCM 15/30/45/60
en(us)
59 / 84
BeWa OLFCM 15-60 3595923a en-us 2018-06-28 .docx
2018-06-28
Red
ERROR
Sensor A is causing
a moderate fault.
Switch the SMU off.
If the error occurs
again, check sensor A
(use HNG 3000 as an
aid)
3
Red
ERROR
Sensor A is causing
a major fault.
Check sensor A (use
HMG3000 as an aid)
4
Red
OUT
RNGE
AS1000 ≤ firmware
V2.04:
The sensor at
sensor interface B is
outside of the
measurement range.
Wait for a few more
measurement cycles.
2
Green
AS1000 ≥ firmware
V2.10:
The sensor at
sensor interface B is
outside of the
measurement range
or has a short circuit
at the sensor.
Wait for a few more
measurement cycles.
Dewater the fluid in
the saturated range.
Check the sensor
outside the fluid or
with the calibration
and adjustment set
(part no. 3122629)
2