Status messages – SMU1200 (optional)
OLFCM 15/30/45/60
en(us)
58 / 84
BeWa OLFCM 15-60 3595923a en-us 2018-06-28 .docx
2018-06-28
Status messages – SMU1200 (optional)
The SMU can take on the following status:
LED
Display
flashing code
Status
To do
Status
-
SMU
no digits displayed
no function
Check the power
supply to the SMU.
Contact the HYDAC
service department.
-
Green
SMU ready for
operation
You can make further
measurements.
0
Red
NoSENS
A sensor is
connected to sensor
interface A.
This is not
recognized.
Check sensor
interface A – is an
MCS1000 or a
CS1000 connected?
Check the connection
cables between the
sensor and the SMU.
Check the sensor bus
address. The bus
address must be
different to SENS B.
Switch the SMU off
and then on again.
If the fault recurs,
contact HYDAC
service department.
3
Green
NoSENS
No sensor is
connected.
This display will go
out after 10
seconds.
Connect a sensor to
sensor interface A.
Switch the SMU off
and then on again.
0