Status messages – CS1000 (optionally with display)
OLFCM 15/30/45/60
en(us)
67 / 84
BeWa OLFCM 15-60 3595923a en-us 2018-06-28 .docx
2018-06-28
LED
Flash code /
display
CS1000 Status
To do
Error
no.
Analogue output
Switching output
Flow status on the
digital output
Red
-
System error
Reset the unit. (To
do this, disconnect
the CS from the
voltage supply) or
contact HYDAC.
-40…-69
4.1 mA / 2.05 V*
open
-
Red
-
Error during
automatic
adjustment
Reset the unit. (To
do this, disconnect
the CS from the
voltage supply)
/ check the flow
rate or contact
HYDAC.
-70
4.1 mA / 2.05 V*
open
-
Red
-
Error measuring
cell LED
Reset the unit. (To
do this, disconnect
the CS from the
voltage supply)
/ check the flow
rate or contact
HYDAC.
-100
4.1 mA / 2.05 V*
open
-
* Is not valid for HDA 5500 output signal