HMG 3010
Page 46
Edition 2013-03-26 - V04 R01
Part No.: 669855
Event: Leaving measurement window
This trigger event is very similar to
Specific
value
. You can specify a
Lower Limit
and
an
Upper Limit
. The trigger is activated
when the measured value enters the meas-
urement window. However, recording is not
started until the measured value leaves the
measurement window in a rising or falling
direction.
Note:
Sub-channel setting only required for SMART sen-
sors (see Section 6.4) and CAN function (see Sec-
tion 6.5).
Event: Timed
The event setting using a time function
(
Timed
) is useful for measurements which
are to be repeated at specific intervals. In
the Settings menu you can specify the num-
ber of
Repetitions
(max. 30,000), the
Inter-
val
and the
Starting time
. In the
Start
item
you can specify whether this event is to be
activated immediately (
Start immediately
)
or not until after a certain
Start date
and
Start time
.
In the
– Event: Timed –
example shown on the right, a measurement
is started 10 times every 30 minutes. The
trigger becomes active for the first time on
21 October 2006 at 14:20.
Up to 4 events can be specified for recording
triggered measurements. After making each
of the settings, return to the
Trigger Set-
tings
menu via
Back
. The menu now pro-
vides an overview of all the events and set-
tings for you to check.