![HYDAC ELECTRONIC HMG 3010 User Manual Download Page 40](http://html.mh-extra.com/html/hydac-electronic/hmg-3010/hmg-3010_user-manual_2191361040.webp)
HMG 3010
Page 40
Edition 2013-03-26 - V04 R01
Part No.: 669855
The sampling rate which can be set is also
dependent on the active measurement
channels:
•
0.1 ms max. 2 analogue input channels
•
0.2 ms max. 4 analogue input channels
•
0.5 ms all 10 input channels
•
1.0 ms for SMART sensors
Recording time
A drop-down list appears when this item is
selected. Using the arrow keys (
▲
) and (
▼
)
you can select a recording time of between 1
second and 24 hours, followed by
OK
.
Similar to when changing the Sampling rate,
information is shown at the bottom of the
display pertaining to the recorded data
showing you the result of your setting.
Recording option
The options available here are
End when
time has elapsed
and
Record continuous-
ly until terminated
.
End when time has elapsed
means that
recording is discontinued when the time en-
tered in
Recording time
has elapsed.
Record continuously until terminated
means that the oldest measured values to
be deleted when new measured values are
added once the time entered in
Recording
time
has elapsed. The time period entered
in
Recording time
remains as the "past" in
relation to the actual time in the measured
value memory until the
Stop
command is
given (
Stop
+
OK
or actuate the
Esc
key).