Start-Stop Measurement
The
Start-Stop Measurement
option allows
you to start a recording when triggered by a
defined condition and to stop it again when
triggered by a second defined condition.
If the stop condition does not occur within
the pre-set recording time, then the record-
ing will stop after the specified time has
elapsed.
Example of a Start-Stop Measurement:
Under
Trigger 1 (Start)
and
Trigger 2
(Stop)
select the relevant start and stop
condition and confirm with
OK
.
Under
Recording time,
define the time at
which the measurement must stop automati-
cally if the Trigger-Condition 2 (Stop) has not
occurred by then.
With
Automatic mode
you define whether
you wish to make just one measurement
(
Deactivated
) or whether, after recording
one measurement curve, another curve is to
be recorded using the same settings (
Acti-
vated
).
Zoomed area
With the "zoomed area", an additional dis-
play area can be created during the record-
ing in the graphic mode.
In other words, with the
Magnifier
setting,
the x-axis is scaled to the time span preset in
this menu point (5 seconds / 10 seconds /
...).
(The setting options for the zoomed area are
dependent on the preset sampling rate and
the specified recording time.)