142
8.3.
Automatic Calibration of Edger
You can execute size calibration and feeler calibration with standard edging Jig.
8.3.1.
Size Calibration
①
Select the ‘Size’ item.
②
Remove the lens adaptor holder and replace it with the standard edging jig.
③
Press the [Start] button to start calibration.
④
You must save the calibration data after finishing the calibration successfully.
8.3.2.
Feeler Calibration
①
Select the ‘Feeler’ item.
②
Rest steps are same with Size Calibration steps.
•
The size calibration data is prerequisite in every type of processing, so size calibration
must be executed above all.
•
It is recommended to execute one calibration at a time.
•
If the “Check the hall sensor contact” message is displayed while using the edger, you
can calibrate the sensor value with the ‘Hall Sensor Contact Calibration’ function.
Calibration function has influence on the edging quality so ask advice of the service technician
of HUVITZ or the technician authorized by HUVITZ if necessary.
!
NOTE
!
CAUTION