140
8.1.1.
Stroke Calibration
Procedures for stroke calibration.
①
Press the ‘Menu’ button on the main screen.
②
Press the ‘Tracer’ button on the menu screen.
③
Select ‘Maintenance(Calibration)’ tab on the screen.
④
Pull the Gripping bars to the maximum.
⑤
Press ‘Stroke’ button to start calibration.
8.1.2.
Frame Calibration
Procedures for frame calibration
①
Press the ‘Menu’ button on the main screen.
②
Press the ‘Tracer’ button on the menu screen.
③
Select ‘Maintenance(Calibration)’ tab on the screen.
④
Load the Standard Frame.
⑤
Press ‘Frame’ button to start calibration.
8.1.3.
Pattern Calibration
Procedures for pattern calibration.
①
Press the ‘Menu’ button on the main screen.
②
Press the ‘Tracer’ button on the menu screen.
③
Select ‘Maintenance(Calibration)’ tab on the screen.
④
Load the Standard Pattern.
⑤
Press ‘Pattern’ button to start calibration.
•
The result of Automatic Calibration will be automatically saved.
•
Be sure to check the related information before executing the calibration work because
you cannot return to the previous state.
•
Press the ‘Tracer Stop’ button on the bottom of the menu screen to stop the calibration.
!
NOTE