Shutting down
OPERATION MANUAL
Chapter 8
Unistat® T300 Series
Liability for errors and misprints excluded.
V1.2.0en/12.05.15//14.06
86
8.2
Switch-off
PROCEDURE
>Mains isolator<
[36] set to “0”.
Disconnect the temperature control unit from the power supply.
8.3
Draining-off the cooling water
This section must be observed when using water-cooled temperature control units.
8.3.1
Draining process
Pressurized cooling water connections
RISK OF INJURY
Wear your personnel protective equipment (e.g. safety goggles).
Carefully open the cooling water connection. Open slowly (1-2 signal edges) and drain the
cooling water slowly.
Building side shut-off valves are not closed
DAMAGE BY ROOM FLOODING
Close the building’s shut-off valves in the cooling water supply and return lines.
For table-top models, place a collection container below the
>Cooling water drain<
[15] (see
wiring diagram page 89, section
).
PROCEDURE
Close the building’s shut-off valves in the cooling water supply and return lines.
Place a collecting container below the
>Cooling water outlet<
[14] and
>Cooling water drain<
[15].
Undo the screw cap on the
>Cooling water drain<
[15]. The cooling water will begin to drain out.
Allow all the cooling water to drain out to prevent the risk of freezing during transport and stor-
age!
8.4
Packing
Use the original packaging wherever possible! Further information can be found on page 19 in sec-
tion
.