Setup mode
OPERATION MANUAL
Chapter 4
Unistat® T300 Series
Liability for errors and misprints excluded.
V1.2.0en/12.05.15//14.06
50
turbance influence. Please note the specifications on page 30 in section
.
The adjustment behavior can be selected at any time without reactivating the “Find control pa-
rameters”.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “TAC/Manual”.
Tap on the sub-category “Configuration auto”.
Tap on the sub-category “Control Dynamics”.
Choose between the dialog entries “Fast, small overshoot” and “Without overshoot”.
Confirm your choice by tapping on “OK”.
Internal, aperiodic
temperature control
Internal, dynamic
temperature control
with possible over-
shooting of the tem-
perature