Commissioning
OPERATION MANUAL
Chapter 2
Unistat® T300 Series
Liability for errors and misprints excluded.
V1.2.0en/12.05.15//14.06
20
▪
Transient surges, as would normally occur in the power supply system
▪
Applicable degree of soiling: 2.
▪
Surge category II.
Side of the
temperature
control unit
Clearance to the temperature control unit in cm
Air cooling
Water cooling
[A1]
Top
Air outlet on top of unit: free standing
–
[A2]
Top
can be located under a bench
can be located under a bench
[B]
Left
min. 20
min. 10
[C]
Right
min. 20
min. 10
[D]
Front
min. 20
min. 10
[E]
Rear
min. 20
min. 20
Side of the
temperature
control unit
Clearance to the temperature control unit in cm (for operation in a bath)
Air cooling
Water cooling
[A1]
Top
Air outlet on top of unit: free standing
–
[A2]
Top
can be located under a bench
can be located under a bench
[B]
Left
min. 20
min. 20
[C]
Right
min. 20
min. 20
[D]
Front
min. 20
min. 20
[E]
Rear
min. 20
min. 20
Wall clearance to
temperature control
unit