Commissioning
Chapter 2
OPERATION MANUAL
V1.2.0en/12.05.15//14.06
Liability for errors and misprints excluded.
Unistat® T300 Series
19
2
Commissioning
2.1
In-plant transport
Temperature control unit transported in a horizontal position
DAMAGE TO THE COMPRESSOR
Only transport the temperature control unit in an upright position.
▪
If fitted, use the lugs located on the top of the temperature control unit for transportation. Do not
transport the temperature control unit alone and without aids.
▪
Protect the temperature control unit from transport damage.
2.2
Unpacking
Starting up a damaged temperature control unit
MORTAL DANGER FROM ELECTRIC SHOCK
Do not operate a damaged temperature control unit.
Please contact the Customer Support. The telephone number can be found on page 88, section
»Phone number and company address«
.
PROCEDURE
Check for damage to the packaging. Damage can indicate property damage to the temperature
control unit.
Check for any transport damage when unpacking the temperature control unit.
Always contact your forwarding agent regarding the settlement of claims.
2.3
Ambient conditions
Unsuitable ambient conditions/unsuitable installation
SERIOUS INJURY DUE TO CRUSHING
Comply with the requirements under sections
and
.
Make sure there is adequate fresh air available at the site for the circulation pump and the com-
pressors. The warm exhaust air must be able to escape upwards unhindered.
Free-standing model
For the connection data, see the data sheet (from page 89 in section
).
Use of the temperature control unit is permitted only under normal ambient conditions in accord-
ance with DIN EN 61010-1:2001:
▪
Use only indoors.
▪
Installation elevation up to 2000 meters above sea level .
▪
Maintain wall and ceiling clearance for adequate air exchange (dissipation of waste heat, supply
of fresh air for the temperature control unit and work area). Ensure adequate floor clearance for
air-cooled temperature control units. Do not operate this temperature control unit from within
the box or with an inadequately dimensioned bath. This inhibits the air exchange.
▪
Ambient temperature values are provided on the technical data sheet; to ensure trouble-free
operation, compliance with the ambient conditions is mandatory.
▪
Relative humidity max 80% to 32 °C and 40 °C decreasing linearly to 50%.
▪
Short distance to supply connections.
▪
The temperature control unit must not be installed so as to hinder or prevent access to the isola-
tor (to the power supply).
▪
Magnitude of the power supply fluctuations: see data sheet from page 89 in section