Tap on the “Start” touchbutton.
Confirm the start of temperature control by tapping on “OK”.
The correct selection will be displayed graphically and temperature control will start immediate-
ly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. Then the dis-
play will return to the “Home” screen again. Try starting the thermoregulation again.
Tap on the keypad symbol next to “T
Setpoint
”.
Enter the new setpoint of 35 °C, using the number keypad that appears.
Confirm your choice by tapping on “OK”.
Open the upstream cooling water hand valve.
Finish the thermoregulation once the flow temperature has reached 40 °C.
Close the upstream cooling water hand valve.
4.3.2
Cooling water connection II
Valid for Unistat T305w HT, T320w HT, T330w HT, T340w HT
Use of unsuitable/defective hoses and/or hose connections
INJURIES
Thermal fluid
Use appropriate hoses and/or hose connections.
Check periodically for leaks and the quality of the hose and hose connections and take suitable
measures (replace) as required.
Isolate and protect temperature control hoses against contact/mechanical load.
Cooling water
Reinforced hoses must be used to satisfy tougher safety requirements.
Shut off the cooling water supply to the temperature control unit even for shorter downtimes
(e.g. overnight).
The T3xxw HT series is equipped with a cooling water connection with controlled cooling water
consumption. These temperature control units not only enable the implementation of positive
temperature gradients (e.g. setpoint step-
changes from 20 °C to 150 °C) but also negative temper
a-
ture gradients (e.g. setpoint step-
changes from 150 °C to 20 °C). The T300w HT series is equipped
with a cooling capacity controller, which regulates the water outlet temperature to a level at which
calcification usually cannot take place (max. 60 °C). Activation of
the cooling water release (as in the
above Section
) is not necessary!
PROCEDURE
Observe the permissible water entry temperature, see data sheet from page 89 in Section
.
Observe the permissible water pressure, see data sheet from page 89 in Section
.
Only use suitable and
temperature-stable
hoses (e.g. reinforced hoses) and screw connections.
4.4
Filling, venting, degassing and draining
The illustration “Wiring diagram” can be found on page 89 in section
.
4.4.1
Externally closed application
Operating an externally closed system with >Isolating sleeve< [27] installed
BURSTING DUE TO OVERPRESSURE DURING TEMPERATURE CONTROL.
Shut down the temperature control unit immediately.
Allow the thermal fluid to cool to room temperature.
Prior to starting up the temperature control unit with an externally closed system, remove the
>Isolating sleeve<
[27] (see page 24, in section
»Mounting/removing the isolating sleeve (op-
).
Open all valves, in case of standard devices the
>Venting valve expansion vessel<
[21] by turn-
ing them anti-
clockwise (turn 90° left as far as it will go).