SOLAR300 - SOLAR300N
EN - 5
1.3. DURING
USE
Please read carefully the below points:
CAUTION
Should you fail to keep to the prescribed instructions you could damage
the instrument and/or its components or endanger your safety
Do not touch any unused terminal when the meter is connected to the
circuit on test
When measuring current, other currents located near the leads may affect
the measuring accuracy
When measuring current, always position the wire in the very middle of the
jaws in order to obtain the highest accuracy
A measured value remains constant if the "
HOLD
" function is active.
Should you notice that the measured value remains unchanged, disable
the “
HOLD
” function
1.4. AFTER
USE
After use, turn off the instrument by pressing
ON/OFF
key for a few seconds.
If you expect not to use the instrument for a long time please keep to the storage
instructions described at § 3.3.
1.5. MEASURING
(OVERVOLTAGE) CATEGORIES DEFINITIONS
IEC/EN61010-1: Safety requirements for electrical equipment for measurement, control
and laboratory use, Part 1: General requirements, gives a definition of measuring
category, usually called overvoltage category. § 6.7.4: Measuring circuits:
(OMISSIS)
Circuits are divided into the following measurement categories:
Measurement category IV
is for measurements performed at the source of the low-
voltage installation
Examples are electricity meters and measurements on primary overcurrent protection
devices and ripple control units
Measurement category III
is for measurements performed in the building installation.
Examples are measurements on distribution boards, circuit breakers, wiring, including
cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and
equipment for industrial use and some other equipment, for example, stationary
motors with permanent connection to fixed installation
Measurement category II
is for measurements performed on circuits directly
connected to the low voltage installation
Examples are measurements on household appliances, portable tools and similar
equipment
Measurement category I
is for measurements performed on circuits not directly
connected to MAINS
Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient stresses are
variable; for that reason, the norm requires that the transient withstand capability of the
equipment is made known to the user