SOLAR300 - SOLAR300N
EN - 64
5.4.2.12. Unbalance
The selection of this option include on selected parameters list for recording the value of
NEG%
and
ZERO%
s which are index of unbalance of input voltage signal respective to
negative tern and zero term (see § 10.8). “Unbalance” is not shown for single phase
systems.
5.4.2.13. Spike
This option permits to set the control parameters relative to recording of voltage fast
transients (spike) events (see § 10.9) with 5
s resolution which is completely
independent
from periodic analysis (regulated by integration period). The below screen is shown by
meter:
Fig. 96: Spike setting screen
Press
F4
key (alternatively touch
MODIFY
at display) for setting the parameter for
detection and recording of voltage spikes. In particular is possible to set:
The value of voltage threshold to detection and recording the voltage spike event from
meter. This value s selectable from
100V
and
5000V
value.
Fig. 97: Spike screen - setting control parameter
1. Move the cursor using arrows keys on field relative to “
Spike
” , marked it with blue
background.
2. Use up or down arrow keys or press
F3
or
F4
keys (alternatively touch the
MOD(+)
or
MOD(-)
s at display) to set the correspondent value. Pressing and holding of these
keys permit a rapid values setting, while the single pressure increase or decrease on
only unit the value.
3. Press
SAVE
or
ENTER
keys (or the smart icon
) to save each performed setting.