3-13
Troubleshooting
Command Error
-113
Undefined header
The header is syntactically correct, but it is undefined for this
specific device. For example,
*XYZ
is not defined for any
device.
-114
Header suffix out of range
The value of a header suffix attached to a program mnemonic
makes the header invalid.
-120
Numeric data error
This error, as well as errors -121 through -129, are generated
when parsing a data element which appears to be numeric,
including non-decimal numeric types. This particular error
message is used if the device cannot detect a more specific
error.
-121
Invalid character in number
An invalid character for the data type being parsed was
encountered. For example, an alpha in a decimal numeric or a
“9” in octal data.
-123
Exponent too large
The magnitude of an exponent was greater than 32000 (see
IEEE 488.2, 7.7.2.4.1).
-124
Too many digits
The mantissa of a decimal-numeric data element contained
more than 255 digits excluding leading zeros (see IEEE 488.2,
7.7.2.4.1).
-128
Numeric data not allowed
A legal numeric data element was received, but the device does
not accept one in this position for the header.
-130
Suffix error
This error, as well as errors -131 through -139, are generated
when parsing a suffix. This particular error message is used if
the device cannot detect a more specific error.
Summary of Contents for ESG-D Series
Page 4: ...ii ...
Page 34: ...Contents 30 ...
Page 46: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Page 138: ...5 2 Hardkey and Softkey Reference ...
Page 164: ...5 28 Hardkey and Softkey Reference Amplitude ...
Page 166: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Page 168: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Page 188: ...5 52 Hardkey and Softkey Reference FM ...
Page 196: ...5 60 Hardkey and Softkey Reference Frequency ...
Page 198: ...5 62 Hardkey and Softkey Reference Help ...
Page 200: ...5 64 Hardkey and Softkey Reference Hold ...
Page 202: ...5 66 Hardkey and Softkey Reference Incr Set ...
Page 236: ...5 100 Hardkey and Softkey Reference LF Out ...
Page 238: ...5 102 Hardkey and Softkey Reference Local ...
Page 240: ...5 104 Hardkey and Softkey Reference Mod On Off ...
Page 242: ...5 106 Hardkey and Softkey Reference Mode ...
Page 336: ...5 200 Hardkey and Softkey Reference Mode GSM ...
Page 422: ...5 286 Hardkey and Softkey Reference Mode PDC ...
Page 466: ...5 330 Hardkey and Softkey Reference Mode PHS ...
Page 532: ...5 396 Hardkey and Softkey Reference Numeric Keypad ...
Page 554: ...5 418 Hardkey and Softkey Reference Power Switch ...
Page 564: ...5 428 Hardkey and Softkey Reference Preset ...
Page 570: ...5 434 Hardkey and Softkey Reference Pulse ...
Page 574: ...5 438 Hardkey and Softkey Reference Recall ...
Page 576: ...5 440 Hardkey and Softkey Reference Return ...
Page 578: ...5 442 Hardkey and Softkey Reference RF On Off ...
Page 584: ...5 448 Hardkey and Softkey Reference Save ...
Page 604: ...5 468 Hardkey and Softkey Reference Sweep List ...
Page 666: ...7 20 Operation Userfile Applications ...
Page 675: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...