2-3
Using Functions
Setting Frequency and Power Levels
7. You can also adjust the RF frequency using the front panel knob. As long as frequency
is the active function (the frequency is displayed in the active entry area), the front panel
knob will increase and decrease the RF frequency. Use the front panel knob to adjust
the frequency back to 700 MHz.
To Set a Frequency Reference and a Frequency Offset
The following procedure sets the RF output frequency as a reference frequency to
which all other frequency parameters are relative. The frequency initially shown
on the display will be 0 Hz (the frequency output by the hardware minus the
reference frequency.) Although the display changes, the frequency output does not
change. Any subsequent frequency changes are shown as incremental or
decremental to 0 Hz.
1. Press
Preset
to return the signal generator to the factory-defined instrument state.
2. Set the RF frequency to 700 MHz.
3. Turn on frequency reference mode and set the current output frequency (700 MHz) as
the reference value. Press
Freq
,
Freq Ref Set
. The frequency displayed is 0 Hz (the
frequency output by the hardware, 700 MHz, minus the reference value, 700 MHz).
Notice that the
REF
indicator is turned on and the
Freq Ref
softkey has toggled to
On
.
4. Set the
RF On/Off
key to
On
. Notice that the display annunciator has changed from
RF OFF
to
RF ON
. The RF frequency at the RF OUTPUT connector is 700 MHz.
5. Increment the output frequency by 1 MHz. Press the up arrow key. The frequency
display changes to show 1 MHz (the frequency output by the hardware,
700 MHz + 1 MHz, minus the reference frequency, 700 MHz) and the output
frequency changes to 701 MHz.
6. Enter a 1 MHz offset. Press the
Freq Offset
softkey and enter 1 MHz using the numeric
keypad and pressing the
MHz
terminator softkey. The frequency display shows 2 MHz
(the frequency output by the hardware, 701 MHz, minus the reference frequency,
700 MHz, plus the offset, 1 MHz.) Notice that the
OFFSET
indicator is turned on. The
frequency at the RF OUTPUT connector is still 701 MHz.
Summary of Contents for ESG-D Series
Page 4: ...ii ...
Page 34: ...Contents 30 ...
Page 46: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Page 138: ...5 2 Hardkey and Softkey Reference ...
Page 164: ...5 28 Hardkey and Softkey Reference Amplitude ...
Page 166: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Page 168: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Page 188: ...5 52 Hardkey and Softkey Reference FM ...
Page 196: ...5 60 Hardkey and Softkey Reference Frequency ...
Page 198: ...5 62 Hardkey and Softkey Reference Help ...
Page 200: ...5 64 Hardkey and Softkey Reference Hold ...
Page 202: ...5 66 Hardkey and Softkey Reference Incr Set ...
Page 236: ...5 100 Hardkey and Softkey Reference LF Out ...
Page 238: ...5 102 Hardkey and Softkey Reference Local ...
Page 240: ...5 104 Hardkey and Softkey Reference Mod On Off ...
Page 242: ...5 106 Hardkey and Softkey Reference Mode ...
Page 336: ...5 200 Hardkey and Softkey Reference Mode GSM ...
Page 422: ...5 286 Hardkey and Softkey Reference Mode PDC ...
Page 466: ...5 330 Hardkey and Softkey Reference Mode PHS ...
Page 532: ...5 396 Hardkey and Softkey Reference Numeric Keypad ...
Page 554: ...5 418 Hardkey and Softkey Reference Power Switch ...
Page 564: ...5 428 Hardkey and Softkey Reference Preset ...
Page 570: ...5 434 Hardkey and Softkey Reference Pulse ...
Page 574: ...5 438 Hardkey and Softkey Reference Recall ...
Page 576: ...5 440 Hardkey and Softkey Reference Return ...
Page 578: ...5 442 Hardkey and Softkey Reference RF On Off ...
Page 584: ...5 448 Hardkey and Softkey Reference Save ...
Page 604: ...5 468 Hardkey and Softkey Reference Sweep List ...
Page 666: ...7 20 Operation Userfile Applications ...
Page 675: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...