Hardkey and Softkey Reference
Recall
Delete Seq[n] Reg[nn]
Press this softkey to delete the contents of a specified register in a specified
sequence. Do not press this softkey until you have selected the correct sequence
and register using the
Select Seq
and
Recall Reg
softkeys. For example, if you
want to delete the contents of register 5 in sequence 4, press
Select Seq
. The
currently selected sequence is displayed in the active entry area. Change the
number to 4 using the up and down arrow keys, the front panel knob, or enter a 4
using the numeric keypad and press the
Enter
terminator softkey. Sequence 4 is
now displayed in all softkeys which specify a sequence in this menu. Now select
register 5. (Remember, you must first have saved an instrument state to this
register in sequence 4.) Press
Recall Reg
. The recall register number becomes the
active function. Enter the number 5 using the numeric keypad and press the
Enter
terminator softkey. The instrument state is restored from this register and register
5 is now displayed in all softkeys which specify a register number in this menu.
Now press
Delete Seq [4] Reg[05]
and register 5, stored in sequence 4, will be
deleted. The instrument state that you had recalled from this register remains,
however, until you change it. If you deleted this register in error, you can
immediately resave it using the
Save Seq[n] Reg[n]
softkey in the
Save
menu.
NOTE:
This softkey has exactly the same function as the softkey with the identical name in the front
panel
Save
key menu.
Softkey Location: Press
Recall
,
Delete Seq[n] Reg[nn]
Recall Reg
Press this softkey to recall an instrument state from a saved register. Recalling a
register also selects that register number for all softkeys which specify a register
number in this menu. For example, if you want to select register 5, press
Recall Reg
. (Remember, you must first have saved an instrument state to this
register in the current sequence identified by the
Select Seq
softkey. If there is
nothing saved, you cannot recall it.) The recall register number becomes the active
function. Enter the number 5 using the numeric keypad and press the
Enter
terminator softkey. The instrument state is restored from this register and register
5 is now displayed in all softkeys which specify a register in this menu. You can
also use the front panel knob and the up and down arrow keys to scroll through all
saved registers for the selected sequence, recalling each instrument state.
Softkey Location: Press
Recall
,
Recall Reg
IEEE 488.2 Common Commands:
*RCL <reg>[, <seq>]
Summary of Contents for ESG-D Series
Page 4: ...ii ...
Page 34: ...Contents 30 ...
Page 46: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Page 138: ...5 2 Hardkey and Softkey Reference ...
Page 164: ...5 28 Hardkey and Softkey Reference Amplitude ...
Page 166: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Page 168: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Page 188: ...5 52 Hardkey and Softkey Reference FM ...
Page 196: ...5 60 Hardkey and Softkey Reference Frequency ...
Page 198: ...5 62 Hardkey and Softkey Reference Help ...
Page 200: ...5 64 Hardkey and Softkey Reference Hold ...
Page 202: ...5 66 Hardkey and Softkey Reference Incr Set ...
Page 236: ...5 100 Hardkey and Softkey Reference LF Out ...
Page 238: ...5 102 Hardkey and Softkey Reference Local ...
Page 240: ...5 104 Hardkey and Softkey Reference Mod On Off ...
Page 242: ...5 106 Hardkey and Softkey Reference Mode ...
Page 336: ...5 200 Hardkey and Softkey Reference Mode GSM ...
Page 422: ...5 286 Hardkey and Softkey Reference Mode PDC ...
Page 466: ...5 330 Hardkey and Softkey Reference Mode PHS ...
Page 532: ...5 396 Hardkey and Softkey Reference Numeric Keypad ...
Page 554: ...5 418 Hardkey and Softkey Reference Power Switch ...
Page 564: ...5 428 Hardkey and Softkey Reference Preset ...
Page 570: ...5 434 Hardkey and Softkey Reference Pulse ...
Page 574: ...5 438 Hardkey and Softkey Reference Recall ...
Page 576: ...5 440 Hardkey and Softkey Reference Return ...
Page 578: ...5 442 Hardkey and Softkey Reference RF On Off ...
Page 584: ...5 448 Hardkey and Softkey Reference Save ...
Page 604: ...5 468 Hardkey and Softkey Reference Sweep List ...
Page 666: ...7 20 Operation Userfile Applications ...
Page 675: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...