5-456
Hardkey and Softkey Reference
Sweep/List
Freq Start
Press this softkey to set the frequency of the first point in the sweep. You can set
the value anywhere in the range of the signal generator’s specified output
frequency. Press
Freq Start
and frequency becomes the active function for the first
point in the sweep. To enter the value, rotate the front panel knob until the desired
value is displayed, use the up and down arrow keys, or enter the value using the
numeric keypad and press the
GHz
,
MHz
,
kHz
, or
Hz
terminator softkey.
Softkey Location: Press
Sweep/List
,
Configure Step Sweep
,
Freq Start
Status after Normal Preset: Maximum specified frequency
SCPI Commands:
[:SOURce]:FREQuency:STARt <val><unit>
[:SOURce]:FREQuency:STARt?
Freq Stop
Press this softkey to set the frequency of the last point in the sweep. You can set
the value anywhere in the range of the signal generator’s specified output
frequency. Press
Freq Stop
and frequency becomes the active function for the last
point in the sweep. To enter the value, rotate the front panel knob until the desired
value is displayed, use the up and down arrow keys, or enter the value using the
numeric keypad and press the
GHz
,
MHz
,
kHz
, or
Hz
terminator softkey.
Softkey Location: Press
Sweep/List
,
Configure Step Sweep
,
Freq Stop
Status after Normal Preset: Maximum specified frequency
SCPI Commands:
[:SOURce]:FREQuency:STOP <val><unit>
[:SOURce]:FREQuency:STOP?
Goto Bottom Row
Press this softkey to move the selection bar to the bottom row of list sweep points
in the Goto Row menu that is accessed from the Configure List Sweep menu. If
you are in the Goto Row menu that is accessed from the Load/Store menu, the
selection bar moves to the bottom row of the catalog of list files.
Softkey Location: Press
Sweep/List
,
Configure List Sweep
,
Goto Row
,
Goto Bottom Row
or press
Sweep/List
,
Configure List Sweep
,
More (1 of 2),
Load/Store
,
Goto Row
,
Goto Bottom Row
Summary of Contents for ESG-D Series
Page 4: ...ii ...
Page 34: ...Contents 30 ...
Page 46: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Page 138: ...5 2 Hardkey and Softkey Reference ...
Page 164: ...5 28 Hardkey and Softkey Reference Amplitude ...
Page 166: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Page 168: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Page 188: ...5 52 Hardkey and Softkey Reference FM ...
Page 196: ...5 60 Hardkey and Softkey Reference Frequency ...
Page 198: ...5 62 Hardkey and Softkey Reference Help ...
Page 200: ...5 64 Hardkey and Softkey Reference Hold ...
Page 202: ...5 66 Hardkey and Softkey Reference Incr Set ...
Page 236: ...5 100 Hardkey and Softkey Reference LF Out ...
Page 238: ...5 102 Hardkey and Softkey Reference Local ...
Page 240: ...5 104 Hardkey and Softkey Reference Mod On Off ...
Page 242: ...5 106 Hardkey and Softkey Reference Mode ...
Page 336: ...5 200 Hardkey and Softkey Reference Mode GSM ...
Page 422: ...5 286 Hardkey and Softkey Reference Mode PDC ...
Page 466: ...5 330 Hardkey and Softkey Reference Mode PHS ...
Page 532: ...5 396 Hardkey and Softkey Reference Numeric Keypad ...
Page 554: ...5 418 Hardkey and Softkey Reference Power Switch ...
Page 564: ...5 428 Hardkey and Softkey Reference Preset ...
Page 570: ...5 434 Hardkey and Softkey Reference Pulse ...
Page 574: ...5 438 Hardkey and Softkey Reference Recall ...
Page 576: ...5 440 Hardkey and Softkey Reference Return ...
Page 578: ...5 442 Hardkey and Softkey Reference RF On Off ...
Page 584: ...5 448 Hardkey and Softkey Reference Save ...
Page 604: ...5 468 Hardkey and Softkey Reference Sweep List ...
Page 666: ...7 20 Operation Userfile Applications ...
Page 675: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...