Using Shock Advisory Mode
4-3
4
Using Shock Advisory Mode
This section provides instructions for using the HP CodeMaster 100 in Shock
Advisory mode. Make certain that you observe all safety precautions noted
throughout this chapter.
Do not use the HP CodeMaster 100 in Shock Advisory mode for:
•children who weigh less than 90 pounds
•patients who have an implanted pacemaker
The advisory algorithm is not designed to detect pediatric cardiac
arrhythmias or handle erratic spiking problems caused by a properly or
improperly functioning pacemaker.
The advisory algorithm does not charge the defibrillator to energy levels
appropriate for pediatric patients.
To use advisory mode, first verify patient condition. Then:
A Select Advisory On.
B
Apply multifunction defib electrodes to the patient.
C Stop patient motion. Press Analyze. If
SHOCK ADVISED
, stand clear.
D When you hear the charge done tone, press both electrode adapter cable Shock
buttons simultaneously.
WA RNI N G
Summary of Contents for Codemaster 100
Page 18: ...Telecom Regulatory Information xviii ...
Page 28: ...xxviii ...
Page 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Page 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Page 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Page 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Page 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Page 178: ...3DFLQJ Using the Pacer 8 8 ...
Page 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Page 234: ...7URXEOHVKRRWLQJ Service 12 22 ...