Using Shock Advisory Mode
4-12
The advisory algorithm needs 7 to 10 seconds of continuous, artifact free ECG data
to decide whether the patient requires defibrillating.
If excessive artifact due to CPR or patient transport is detected, the monitor displays
the messages:
ARTIFACT DETECTED
DO NOT TOUCH PATIENT
.
If the artifact ends within 20 seconds, analysis continues. Otherwise, the following
messages appear on the monitor, and analysis stops. You can restart analysis by
pressing Analyze.
ARTIFACT DETECTED
CANNOT ANALYZE
Pressing the Analyze key during analysis will stop analysis. The message
ANALYSIS STOPPED
will be displayed momentarily, and then the following
message appears:
FOR ANALYSIS,
200J
PRESS ANALYZE
SELECTED
If the multifunction defib electrodes are not properly attached to the patient when
you press Analyze, the following message appears.
FOR ANALYSIS,
200J
ATTACH PADS
SELECTED
The electrodes must be attached to the patient and the electrode adapter cable to start
analysis.
NO TE
NO TE
Summary of Contents for Codemaster 100
Page 18: ...Telecom Regulatory Information xviii ...
Page 28: ...xxviii ...
Page 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Page 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Page 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Page 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Page 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Page 178: ...3DFLQJ Using the Pacer 8 8 ...
Page 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Page 234: ...7URXEOHVKRRWLQJ Service 12 22 ...